A systematic review with meta-analysis of the diagnostic test accuracy of pedicle screw electrical stimulation

Data de publicação: Data Ahead of Print:

Autores da FMUP

  • Manuel António Pereira Gutierres

    Autor

Participantes de fora da FMUP

  • Fonseca, P
  • Goethel, M
  • Vilas-Boas, J.
  • Correia, MV

Unidades de investigação

Abstract

Purpose To provide a systematic review with meta-analysis providing evidence of the current diagnostic test accuracy (DTA) of pedicle screw electrical stimulation. Methods A systematic database search on PubMed, Scopus and Web of Science was performed according to the PRISMA-DTA guidelines, and eligibility criteria applied to reduce the results to: (1) only journal articles reporting electrical stimulation of the pedicle screw head, (2) screw position confirmation by imaging techniques, and (3) enough information allowing the calculation of a 2 x 2 contingency table. Sample characteristics, image confirmation method, electrical current threshold and stimulation results were retrieved and analyzed using according to appropriate DTA analysis methods, and allowing the calculation of specificity, sensitivity for pedicle screws insertion at the lumbar and thoracic levels. Results Lumbar screw stimulation presents a higher sensitivity (0.586 [0.336, 0.798] and specificity (0.984 [0.958, 0.994]) than thoracic screws (sensitivity: 0.270 [0.096; 0.562]; specificity: 0.958 [0.931, 0.975]). The same is observed in terms of the diagnostic odds ratio for lumbar (88.32 [32.136, 242.962]) and thoracic (8.460 [2.139, 33.469]) levels. When performing a sub-group analysis, it is possible to divide the lumbar stimulation threshold as 8 and 10-12 mA, and the thoracic threshold as 6 and 9-12 mA. A threshold of 8 mA at the lumbar level provides higher sensitivity and specificity. Increasing the threshold results in higher specificity but not sensitivity. In fact, at the range of 10-12 mA, the diagnostic validity is too low to confer this technique any robust diagnostic validity. Similarly, at the thoracic level, lower threshold currents are associated with increased sensitivity, but their diagnostic validity is very low. Conclusion Electrical stimulation of the pedicle screw can be used as an adequate diagnostic capability at the lumbar level with a threshold of 8 mA. However, thoracic stimulation is currently not reliable, with very low sensitivity and diagnostic validity at 6 mA or higher.

Dados da publicação

ISSN/ISSNe:
0940-6719, 1432-0932

European Spine Journal  Springer Verlag

Tipo:
Review
Páginas:
1599-1610
Link para outro recurso:
www.scopus.com

Citações Recebidas na Web of Science: 1

Citações Recebidas na Scopus: 4

Documentos

  • Não há documentos

Métricas

Filiações mostrar / ocultar

Keywords

  • Diagnostic accuracy; Pedicle screw; Intraoperative; Neuromonitoring; Electrical stimulation; Spine surgery

Financiamento

Proyectos asociados

Pseudo-Patela Baixa após Artroplastia Total do Joelho: Avaliação Radiológica e Repercussão Clínica

Investigador Principal: Manuel António Pereira Gutierres

Estudo Clínico Académico . 2021

COVID-19 Lockdown: The Impact of Home-based Training Guidance in Prevention of Football Injuries During the Return to Competition

Investigador Principal: Manuel António Pereira Gutierres

Estudo Clínico Académico . 2021

Transferências tendinosas para o tratamento de roturas maciças póstero-superiores da coifa: trapézio inferior vs. grande dorsal

Investigador Principal: Manuel António Pereira Gutierres

Estudo Clínico Académico . 2021

Reverse Shoulder Arthroplasty with (or without) subscapularis repair: does it affect. range of motion and/or articular stability?

Investigador Principal: Manuel António Pereira Gutierres

Estudo Clínico Académico . 2021

Avaliação Isocinética do Músculo Deltoide Após Artroplastia Invertida do Ombro

Investigador Principal: Manuel António Pereira Gutierres

Estudo Clínico Académico . 2020

Citar a publicação

Partilhar a publicação