Pulmonary function electronic monitoring devices - A randomized agreement study
Autores da FMUP
Participantes de fora da FMUP
- Magalhaes, M
- Castel Branco, MG
- Vaz, M
Abstract
Study objectives: To compare in a clinical setting the within-session reproducibility, of two pulmonary function electronic monitoring devices (PiKo-1; Ferraris Respiratory Europe; Hereford, UK; and Spirotel; MIR; Rome, Italy) with one mechanical device (Mini-Wright Peak-Flow Meter; Clement-Clarke International; Harlow, Essex, UK), and to evaluate the accuracy of these devices using as reference an office pneumotachograph. Design, setting, and participants: After detailed instructions, adults without airways diseases and patients with stable asthma attending an outpatient clinic performed four sets of expiratory maneuvers, one set for each device, in a strictly, random order. Each set comprised three maneuvers with 2 to 3 min of rest between them. Measurements: Reproducibility of FEV1 and peak expiratory, flow (PEF) was assessed by a coefficient of variation (CV) and intraclass correlation coefficient (ICC), and accuracy was assessed by ICC and limits of agreement. Results: Of the 38 participants evaluated, 71% were women and 61% had asthma. Ages ranged from 18 to 58 years, and FEV1 ranged from 1.2 to 4.8 L. In all monitoring devices, CV was < 6% and ICC was > 0.94 for the reproducibility of both FEV, and PEF measurements. The accuracy of the PiKo-1 device was better for FEV, (ICC = 0.98) than for PEF (ICC = 0.90). The Spirotel device had similar results for FEV, and PEF (ICC = 0.95). The Mini-Wright device had the lowest accuracy (ICC = 0.87), particularly for PEF values < 500 L/min. Conclusions: These low-cost and easy-to-use electronic monitoring devices showed a very good reproducibility and were in agreement with the pneumotachograph. Therefore, the PiKo-I and Spirotel devices seem adequate for both screening and monitoring. However, prospective studies are still needed to assess their long-term reproducibility and usability and, particularly, the effects on the improvement of respiratory, care.
Dados da publicação
- ISSN/ISSNe:
- 0012-3692, 1931-3543
- Tipo:
- Article
- Páginas:
- 1258-1265
- PubMed:
- 16162716
- Link para outro recurso:
- www.scopus.com
Chest American College of Chest Physicians
Citações Recebidas na Web of Science: 45
Citações Recebidas na Scopus: 58
Documentos
- Não há documentos
Filiações
Filiações não disponíveis
Keywords
- adult; article; asthma; biomedical technology assessment; clinical trial; comparative study; controlled clinical trial; controlled study; correlation coefficient; device; diagnostic accuracy; female; forced expiratory volume; human; lung function; lung function test; major clinical study; male; monitoring; outpatient; pneumotachygraphy; priority journal; quality of life; randomized controlled trial; reproducibility; screening
Citar a publicação
Fonseca J,da Costa Pereira A,Delgado L,Nogueira L,Magalhaes M,Castel MG,Vaz M. Pulmonary function electronic monitoring devices - A randomized agreement study. Chest. 2005. 128. (3):p. 1258-1265.
Portal de investigação