Mock circulatory system for the evaluation of left ventricular assist devices, endoluminal prostheses, and vascular diseases.

Data de publicação: Data Ahead of Print:

Autores da FMUP

  • João De Almeida Lopes Da Fonseca

    Autor

Participantes de fora da FMUP

  • Legendre D
  • Andrade A
  • Biscegli JF
  • Manrique R
  • Guerrino D
  • Prakasan AK
  • Ortiz JP
  • Lucchi JC

Abstract

A new digital computer mock circulatory system has been developed in order to replicate the physiologic and pathophysiologic characteristics of the human cardiovascular system. The computer performs the acquisition of pressure, flow, and temperature in an open loop system. A computer program has been developed in Labview programming environment to evaluate all these physical parameters. The acquisition system was composed of pressure, flow, and temperature sensors and also signal conditioning modules. In this study, some results of flow, cardiac frequencies, pressures, and temperature were evaluated according to physiologic ventricular states. The results were compared with literature data. In further works, performance investigations will be conducted on a ventricular assist device and endoprosthesis. Also, this device should allow for evaluation of several kinds of vascular diseases.

Dados da publicação

ISSN/ISSNe:
1525-1594, 0160-564X

Artificial Organs  Wiley-Blackwell Publishing Ltd

Tipo:
Article
Páginas:
461-467
PubMed:
18422796

Citações Recebidas na Web of Science: 36

Documentos

  • Não há documentos

Métricas

Filiações mostrar / ocultar

Citar a publicação

Partilhar a publicação