Development and evaluation of an algorithm for computer analysis of maternal heart rate during labor

Data de publicação:

Autores da FMUP

  • João Francisco Montenegro De Andrade Lima Bernardes

    Autor

  • Cristina Maria Nogueira Da Costa Santos

    Autor

Participantes de fora da FMUP

  • Pinto, Paula
  • Amorim-Costa, Celia
  • Silva, Maria
  • Ayres-de-Campos, Diogo

Abstract

Background: Maternal heart rate (MHR) recordings are morphologically similar and sometimes coincident with fetal heart rate (FHR) recordings and may be useful for maternal-fetal monitoring if appropriately interpreted. However, similarly to FHR, visual interpretation of MHR features may be poorly reproducible. Methods: A computer algorithm for on-line MHR analysis was developed based on a previously existing version for FHR analysis. Inter-observer and computer-observer agreement and reliability were assessed in 40 one-hour recordings obtained from 20 women during the last 2 h of labor. Agreement and reliability were evaluated for the detection of basal MHR, long-term variability (LTV), accelerations and decelerations, using proportions of agreement (PA) and Kappa statistic (K), with 95% confidence intervals (95% Cl). Changes in MHR characteristics between the first and the second hour of the tracings were also evaluated. Results: There was a statistically significant inter-observer and computer-observer agreement and reliability in estimation of basal MHR, accelerations, decelerations and LTV, with PA values ranging from 0.72 (95% CI: 0.62-0.79) to 1.00 (95% CI: 0.99-1.00), and K values ranging from 0.44(95% CI: 0.28-0.60) to 0.89 (95% CI: 0.82-0.96). Moreover, basal MHR, number of accelerations and LTV were significantly higher in the last hour of labor, when compared to the initial hour. Discussion: The developed algorithm for on-line computer analysis of MHR recordings provided good to excellent computer-observer agreement and reliability. Moreover, it allowed an objective detection of MHR changes associated with labor progression, providing more information about the interpretation of maternal-fetal monitoring during labor. (C) 2014 Elsevier Ltd. All rights reserved.

Dados da publicação

ISSN/ISSNe:
0010-4825, 1879-0534

Computers in Biology and Medicine  Elsevier Ltd.

Tipo:
Article
Páginas:
30-35
PubMed:
24727565
Link para outro recurso:
www.scopus.com

Citações Recebidas na Web of Science: 15

Citações Recebidas na Scopus: 19

Documentos

  • Não há documentos

Métricas

Filiações mostrar / ocultar

Keywords

  • Fetal heart rate; Maternal heart rate; Cardiotocography; Computer analysis; Agreement; Reliability