The endotracheal tube biases the estimates of pulmonary recruitment and overdistension

Data de publicação:

Autores da FMUP

  • Alysson Roncally Silva Carvalho

    Autor

Participantes de fora da FMUP

  • Jandre, Frederico C.
  • Modesto, Felipe Cardozo
  • Antonio Giannella-Neto

Abstract

To assess the impact of the endotracheal tube (ETT) and of different flow waveforms on estimates of alveolar cyclic recruitment (CR) and overdistension (AO). Numerical simulation of the respiratory system plus ETT (inertance L plus a flow-dependent resistance, K-1 and K-2), with the following non-linear equation of motion P-AW(t) = ((K-1 + K-2 center dot vertical bar <(V)over dot>vertical bar) center dot (V)over dot >(t) + L center dot <(V)double over dot > (t)) + Rrs center dot (V)over dot >vertical bar (t) + (E-1 + E-2 center dot V(t)) center dot V(t) + P-0 (P-AW pressure at the airways opening, V volume), under volume-controlled mechanical ventilation. An index %E2 = 100 center dot(E-2 center dot V-T)/( E-1 + E-2 center dot VT) can be calculated where %E-2 > 30% represents AO and %E-2 < 0% represents CR. Parameters were estimated by the least-squares method, either with the complete equation or supressing L, K-2 or both. %E-2 is always underestimated (down to -152 percent points) with incomplete equations of motion. The estimation of %E-2 may be strongly biased in the presence of an ETT excluded from the estimation model.

Dados da publicação

ISSN/ISSNe:
1741-0444, 0140-0118

MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING  Springer Verlag

Tipo:
Article
Páginas:
69-73
PubMed:
17932699
Link para outro recurso:
www.scopus.com

Citações Recebidas na Web of Science: 15

Citações Recebidas na Scopus: 16

Documentos

  • Não há documentos

Métricas

Filiações mostrar / ocultar

Keywords

  • respiratory mechanics; alveolar overdistension; tidal recruitment; model identification; non-linear models

Partilhar a publicação