Cupid fixation for repositioning subluxated intraocular lens

Data de publicação:

Autores da FMUP

  • Fernando Manuel Mendes Falcão Reis

    Autor

Participantes de fora da FMUP

  • Domingues, M
  • Brito, P
  • Falc?o, M.
  • Monteiro, T

Abstract

We describe the cupid fixation technique, which allows safe repositioning of a subluxated intraocular lens (IOL). Under subconjunctival anesthesia, the body of the subluxated IOL is perforated with a 10-0 polypropylene suture on a straight needle. The IOL is then centered and fixated at the sclera overlying the ciliary sulcus; the knot is tied beneath a previously created limbal intrascleral pocket. No intraoperative complications occurred in 24 cases in which the technique was performed, and successful IOL centration was achieved. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned. J Cataract Refract Surg 2011; 37:1571-1575 (C) 2011 ASCRS and ESCRS