Validation of a New Minimally Invasive Intracranial Pressure Monitoring Method by Direct Comparison with an Invasive Technique

Data de publicação:

Autores da FMUP

  • Maria Celeste Pinheiro Dias Ferreira

    Autor

Participantes de fora da FMUP

  • Frigieri Vilela, Gustavo Henrique
  • Cabella, Brenno
  • Mascarenhas, Sergio
  • Czosnyka, Marek
  • Smielewski, Peter
  • Cardim, Danilo Augusto
  • Mascarenhas, Yvonne Maria
  • Wang, Charles Chenwei
  • Andrade, Rodrigo
  • Tanaka, Koji
  • Lopes, Luiza Silva
  • Colli, Benedicto Oscar

Abstract

In this chapter we present in vivo experiments with a new minimally invasive method of monitoring intracranial pressure (ICP). Strain gauge deformation sensors are externally glued onto the exposed skull. The signal from these sensors is amplified, filtered, and sent to a computer with appropriate software for analysis and data storage. Saline infusions into the spinal channel of rats were performed to produce ICP changes, and minimally invasive ICP and direct Codman intraparenchymal ICP were simultaneously acquired in six animals. The similarity between the invasive and minimally invasive methods in response to ICP increase was assessed using Pearson's correlation coefficient. It demonstrated good agreement between the two measures < r > = 0.8 +/- 0.2, with a range of 0.31-0.99.

Dados da publicação

ISBN:
9783319225333

Intracranial Pressure And Brain Monitoring Xv  SPRINGER

Tipo:
Proceedings Paper
Páginas:
97-100
PubMed:
27165885

Citações Recebidas na Web of Science: 20

Documentos

  • Não há documentos

Filiações mostrar / ocultar

Keywords

  • Intracranial pressure (ICP); Medical instrumentation; Minimally invasive system; Monitoring