Establishment of Diagnostic Reference Levels in Portuguese Interventional Radiology departments
Autores da FMUP
Participantes de fora da FMUP
- Pimenta, A
- Ramos, I
- Santos, J
Unidades de investigação
Abstract
Purpose: To establish Portuguese Diagnostic Reference Levels (DRLs), for six body fluoroscopy guided interventional procedures (FGIP). Method: A retrospective study was conducted in five interventional departments most representative of Interventional Radiology (IR) practice. Dose values, in terms of air kerma area product (PKA in Gy.cm2), air kerma at the patient entrance reference point (Ka,r in mGy), and exposure parameters (fluoroscopy time (FT) and number of cine runs) were collected. Examinations were selected per procedure (at least 20), according to the anteroposterior and lateral diameter mean value (+/- 5 cm), measured on previous Computed Tomography (CT) examinations. Results: Data of 489 body FGIP show a large variation on dose values per procedure and per department. National DRLs in terms of PKA were 20.2 Gy.cm2 for Percutaneous transhepatic biliary drainage (PTBD), 98.2 Gy.cm2 for Bronchial artery embolisation (BAE), 247.7 Gy.cm2 for Transarterial chemoembolisation (TACE), 331.6 Gy.cm2 for Inferior epigastric arteries embolisation (IEAE), 312.0 Gy.cm2 for Transjugular intrahepatic portosystemic shunt (TIPS) and 19.3 Gy.cm2 for Endovascular treatment of femoral popliteal arteries (ETFPA). Conclusions: This is the first study reporting Interventional Radiology DRLs in Portugal and we propose preliminary national estimates for the six more common body FGIP. The results of this study will be presented and discussed with all Portuguese IR departments, to promote procedures optimisation.
Dados da publicação
- ISSN/ISSNe:
- 1872-7727, 0720-048X
- Tipo:
- Article
- Páginas:
- -
- Link para outro recurso:
- www.scopus.com
European Journal of Radiology Elsevier Ireland Ltd
Documentos
- Não há documentos
Filiações
Filiações não disponíveis
Keywords
- Interventional Radiology; Diagnostic Reference Levels; Radiation Protection; Dose values
Campos de estudo
Proyectos asociados
Bioavailability of EPA + DHA Enriched Canned Tuna and Its Acute Effects on Cardiovascular Risk Markers, in Healthy Human Volunteers (FUNCTIONALTUNA) - NCT03742492
Investigador Principal: Luís Filipe Ribeiro de Azevedo
Ensaio Clínico Académico (FUNCTIONALTUNA) . CINTESIS . 2019
Health Tecnology Assessment of Nutrition an Food Productd - The Scientific Evidence Behind Health Claims
Investigador Principal: Luís Filipe Ribeiro de Azevedo
Estudo Clínico Académico . 2020
Comparative effectiveness of opioid therapy for the long term management of chronic non cancer pain
Investigador Principal: Luís Filipe Ribeiro de Azevedo
Estudo Clínico Académico . 2020
Citar a publicação
Pimenta A,Azevedo L,Ramos I,Santos J. Establishment of Diagnostic Reference Levels in Portuguese Interventional Radiology departments. Eur. J. Radiol. 2024. 173. 111377. IF:3,300. (2).