Variation in the measurement of DNA damage by comet assay measured by the ECVAG inter-laboratory validation trial

Data de publicação:

Autores da FMUP

  • João Paulo Fernandes Teixeira

    Autor

Participantes de fora da FMUP

  • Forchhammer, L
  • Johansson, C
  • Loft, S
  • Möller, L
  • Godschalk, RWL
  • Langie, SAS
  • Jones, GDD
  • Kwok, RWL
  • Collins, AR
  • Azqueta, A
  • Phillips, DH
  • Sozeri, O
  • Stepnik, M
  • Palus, J
  • Vogel, U
  • Wallin, H
  • Routledge, MN
  • Handforth, C
  • Allione, A
  • Matullo, G
  • Costa, S
  • Riso, P
  • Porrini, M
  • Moller, P

Abstract

The comet assay has become a popular method for the assessment of DNA damage in biomonitoring studies and genetic toxicology. However, few studies have addressed the issue of the noted inter-laboratory variability of DNA damage measured by the comet assay. In this study, 12 laboratories analysed the level of DNA damage in monocyte-derived THP-1 cells by either visual classification or computer-aided image analysis of pre-made slides, coded cryopreserved samples of cells and reference standard cells (calibration curve samples). The reference standard samples were irradiated with ionizing radiation (0-10 Gy) and used to construct a calibration curve to calculate the number of lesions per 10(6) base pair. All laboratories detected dose-response relationships in the coded samples irradiated with ionizing radiation (1.5-7 Gy), but there were overt differences in the level of DNA damage reported by the different laboratories as evidenced by an inter-laboratory coefficient of variation (CV) of 47%. Adjustment of the primary comet assay end points by a calibration curve prepared in each laboratory reduced the CV to 28%, a statistically significant reduction (P < 0.05, Levene's test). A large fraction of the inter-laboratory variation originated from differences in image analysis, whereas the intra-laboratory variation was considerably smaller than the variation between laboratories. In summary, adjustment of primary comet assay results by reference standards reduces inter-laboratory variation in the level of DNA damage measured by the alkaline version of the comet assay.

Dados da publicação

ISSN/ISSNe:
1464-3804, 1464-3804

Mutagenesis  Oxford University Press

Tipo:
Article
Páginas:
113-123
PubMed:
19910383
Link para outro recurso:
www.scopus.com

Citações Recebidas na Web of Science: 141

Citações Recebidas na Scopus: 158

Documentos

  • Não há documentos

Métricas

Filiações mostrar / ocultar

Keywords

  • Calibration; Cells, Cultured; Comet Assay; Cryopreservation; DNA Damage; DNA-Formamidopyrimidine Glycosylase; Humans; Laboratories; Monocytes; Observer Variation; Radiation, Ionizing; Reference Standards; cesium 137; DNA; article; calibration; comet assay; computer program; cryopreservation; DNA damage; DNA strand breakage; dose response; gamma irradiation; genetic toxicology; human; human cell; image analysis; ionizing radiation; laboratory; measurement; monocyte; priority journal; radiation dose; reference value; scoring system; standard; validation study