Twiddler (or Not) Syndrome: Questioning etiology for an uncommon form of hardware malfunction in deep brain stimulation

Data de publicação:

Autores da FMUP

  • Paulo José Campos Linhares Vieira

    Autor

Abstract

Background: Hardware failure or malfunction after deep brain stimulation is an infrequent but costly occurrence with currently available systems. Case Description: The authors present the case of a 65-year-old female patient with predominantly tremoric Parkinsons disease who, 4 months after bilateral subthalamic nucleus stimulation with very good clinical results, began to display signs of recurrent disease and an increasingly smaller response to stimulation. Radiological studies, changes in electrode impedance and surgical findings and results established the diagnosis of Twiddler syndrome. Close patient follow-up, lack of a psychiatric history and physical examination findings were, however, contrary to the previously described causative mechanism. Conclusion: The clinical and radiological setup of Twiddler syndrome must be readily recognized. Its causative mechanism should remain under discussion, and intraoperative technical details may help to explain its occurrence. ? 2014 Silva PA.

Dados da publicação

ISSN/ISSNe:
2152-7806,

Surgical Neurology International  Scientific Scholar LLC

Tipo:
Article
Páginas:
410-

Documentos

  • Não há documentos

Métricas

Filiações

Filiações não disponíveis

Citar a publicação

Partilhar a publicação